Core Lab Equipment: Amersham Typhoon 9200 Imager


Description:

The Typhoon instrument is a variable-mode imager that produces digital images of radioactive, fluorescent, or chemiluminescent samples.  The Typhoon system consists of an instrument, instrument control software, and a variety of accessories.  Image Quant Software includes: FragmeNT Analysis V1.2, Gene Sight V3.2, ImaGene V3.2, Image Quant V5.2, IQTools V2.2, FluorSep V2.2, and Clone Tracker V2.0.

  • The Typhoon scans and processes samples in the following way:
  • Red laser illuminates each one-pixel section of an exposed storage phosphor screen, the phoshphor screen emits blue light.
  • When the green laser (532 nm) or the red laser (633 nm) illuminates each one-pixel section of a fluorescent sample, the fluorochrome emits light with a characteristic spectrum.
  • The chemiluminescent sample emits light with a characteristic spectrum. 
  • Multilabel-excited mode-Creates an image from a sample labeled with up to four fluorescent dyes. The dyes can be excited by one or two of the laser wavelengths.

 Note: EtBr samples are not allowed in the Typhoon, there is an Alpha Imager specifically for imaging these gels.

Accessories:

  • Image Eraser-Erases storage phosphor screens for reuse
  • Storage phosphor screen-General-purpose storage phosphor screen that is 20cm x 25 cm
  • Exposure cassette- Two small cassettes for unmounted screens

Standard set of emission filters:

Emission filter

Description

555-nm band-pass filter (555 BP 20)

Transmits light between 545 nm and 565 nm and has a transmission peak centered at 555 nm.  Use this filter with fluorochromes, such as R6G and HEX.

580-nm band pass filter (580 BP 30)

Transmits light between 565 nm and 625 nm and has a transmission peak centered at 580 nm.  Use this filter with fluorochromes, such as TAMRA and Cy3.

610-nm band-pass filter (610 BP 30)

Transmits light between 595 nm and 625 nm and has a transmission peak centered at 610 nm.  Use this filter with fluorochromes, such as ROX, EtBr, SYPRO Red, and SYPRO Ruby.

670-nm band pass filter (670 BP 30)

Transmits light between 655 and 685 nm and has a transmission peak centered at 670 nm.  Use this filter with fluorochromes, such as Cy5.

526-nm short-pass filter

Transmits light below 526 nm.  Use this filter with fluorochromes, such as fluorescein when scanning with the green laser.  (Note:  When using the blue laser, the 520 BP 40 might provide better results.)

560-nm long-pass filter

Transmits light above 560 nm.  Use this filter with fluorochromes, such as TRITC.

390-nm band-pass filter

This filter is used for storage phosphor scanning and is not normally used for fluorescence scanning.

 

The Typhoon contains three Standard beamsplitters and has a slot for one additional custom beamsplitter. 

Beamsplitter

Description

560-nm Dichroic

Reflects light wavelengths shorter than 560 nm and passes light longer than 560 nm.

580-nm dichroic

Reflects light wavelengths shorter than 580 nm and passes light longer than 580 nm

630-nm dichroic

Reflects light wavelengths shorter than 630 nm and passes light longer than 630 nm

Location:

Core Lab, WRRB 207